This compact gauge can be used for non-destructive coating thickness measurement of non-magnetic coatings, e.g. paint,enamel, chrome on steel, and insulating coatings, e.g. paint and anodizing coatings on non-ferrous metals.
The internal probe can work on both principles, magnetic induction and the eddy currents. The probe can automatically detect the substrates type (Magnetic or not), and calculate the coating thickness and display it fast.
This is an excellent contractors gauge.
NZCS 550 FN Contractor Gauge
FEATURES
1. 128*128 dot matrix LCD display, standard menu operations;
2. Two measure mode: single and continuous;
3. Two group mode: direct (DIR) and general(GEN), readings will be lost when power off in direct mode, and not be lost ingeneral mode. 80 readings can be stored for each group;
4. Zero-point calibration and multi-point calibration (up to 4 points) for each group;
5. User can recall, delete specified readings, or delete group readings;
6. Statistics display: mean, minimum, maximum and standard deviation;
7. Three probe mode: auto, magnetic and eddy current;
8. User can set high or low limit alarm for each group;
9. Power off automatically;
10. USB interface to data transmission;
11. Low battery and error indication
SPECIFICATIONS
1. Measuring principle: Magnetic induction (F-probe) and eddy current (N-probe);
2. Measuring range: 0 to 1300um (0 to 51.2mils);
3. Accuracy: ± (3% of readings+2um);
± (3% of readings+0.078mils);
4. Resolution: 0um~999um (1um), 1000um~1300um (0.01mm);
0mils~39.39mils (0.01mils), 39.4mils~51.2mils (0.1mils);
5. Calibration: One to four-point calibration, zero calibration;
6. Data Group: One direct group (readings not be stored to memory), four general group (readings can be stored), and each
group have individual statistics, alarm settings and calibration;
7. Statistics: No. of readings, mean, minimum, maximum and standard deviation;
8. Units: um, mm and mils;
9. Alarm: User can set the high/low alarm, and alarm icon displayed on LCD when over the limit;
10. Minimum curvature radius: convex 1.5mm(59mils) and concave 25mm(984mils);
11. Minimum measuring size: Diameter 6mm(236mils);
12. Minimum thickness of substrate: F-probe: 0.5mm(0.02"), N-probe: 0.3mm(0.012");
13.Computer interface: Download data via USB interface;
14. Power supply: Two 1.5v AAA battery;
15. Operation temperature: 0℃to 40℃(32℉to 104℉);
16. Storage temperature: -20℃to 70℃(-4℉to 158℉);
17. Size: 110mm*53mm*24mm (4.33"*2.09"*0.94");
18. Weight: 92g (3.24oz);
Corrosion and Materials Engineering Services
For over 11 years we are pleased to offer our on-going support for training and development in the protective coatings and linings industries throughout New Zealand & Australia.
We support the following organisations in our roll.